▼Description
The beamline BL3U is equipped with an in-vacuum undulator composed of 50 periods of 3.8 cm period length.
The emitted photons are monochromatized by the varied-line-spacing plane grating monochromator (VLS-PGM)
designed for various spectroscopic investigations in the soft X-ray range including soft X-ray emission studies.
Three holographically ruled laminar profile plane gratings are designed to cover the photon energy range from
60 to 800 eV. The beamline has two endstations, namely, XES setup and multi-purpose setup. The XES setup is
used for soft X-ray emission spectroscopy. The beam is horizontally focused onto the sample position by a
plane-elliptical mirror, M2X. In the multi-purpose setup, the beam is focused by the toroidal mirror M2. Between
the sample position and M2, the differential pumping is placed. XAS of liquid samples and the application of in
operando observations are performed at the experimental stage of the multi-purpose setup.
Fig. 1. Schematic layout (left) and the photography (right) of the BL3U. The distances along the beam from the
center of the in-vacuum plane undulator are shown in millimeters. S1X and M2X can be replaced with the other
exit slit S1 so that experiments can be carried out at either the XES or the multipurpose endstation. In the XES
setup, the sample is placed 5–10 mm downstream of S1X.
▼Technical Data
Monochromator | Varied-line-spacing plane grating monochromator |
Energy Range | 60-800 eV |
Resolution | E / ΔE > 10 000 |
Experiments | Soft X-ray spectroscopy (XPS, XES, XAS) |
Beam Size (XES Endstation |
Gaussian shape Vertical 5-20 μm; Horizontal 41 μm (FWHM) |